| 1 |
Carl Zeiss IOL Master Opthalmic Biometry Instrument
|
Year of manufacture:- 2002
Software version:- Version option A, version 3.01.0204
To include:- HP Deskjet 940C jet printer, Keysonic ACK-54D keyboard - serial number 103280568NH and a Carl Zeiss test eye
***Our Engineer’s comments:- powers up, self test and boot up ok. Image of test eye obtained of LCD screen. Key and touch pad in good working order. Axial length measurement was made with test eye, and same value was achieved by the IOL master. Data storage and review was also possible. Motorised height adjustment of table is also working ok. System in good condition and functional at the time of test.
|
500 - 800 |
|
| 2 |
Kapps Asslar Colposcope
|
Serial number:- 8567
To include:- DP Medical 404CCD camera - serial number 6020188 and a manual
***Our Engineer’s comments:- Powers up and lamp comes on, brightness control for lamp in good working order. Good magnification with the binocular set, clear optical system and focusing. The attached camera - also fully functional with very good video image output. Tested with a colour monitor. System in very good condition and functional at the time of test.
|
700 - 1000 |
|
| 3 |
Olympus Stack Trolley
|
To include:- Olympus CV240 video system centre - serial number 2925012, Olympus CLVU20 light source - serial number 7132756, Olympus CLVU40 light source - serial number 7804623, Sony monitor, Olympus keyboard and a foot switch.
***Our Engineer’s comments:- Both light sources power up ok. Xenon lamps light up and are fully operational. Lamp life indicator not more than 50 hours. All controls are functional and in good condition. Video processor in good working order, powers up ok, all controls and settings operational. Good image quality. Keyboard and colour monitor are working ok. System is in good condition at the time of test.
|
1000 - 2000 |
|
| 4 |
GE Medical Systems Logiq 5 Expert Ultrasound
|
Serial number:- C3281SU3
Year of manufacture:- 2003
Software Version:- R.3.13
Service Record:- up to Feb 2009
To include:- Sony UP-855MDW b/w printer - serial number 31798, GE 2294517 5.0Mhz probe - serial number 940184YM5, GE 2050357 REV1 2.5Mhz probe - serial number 17179PD3
Our Engineer’s comments:- Powers up and system self test/boot up ok. All controls and display are fully functional. Touch screen and menu navigation are acceptable and in good working order. Both probes are working and in good condition. All tested functions of all available modes are in satisfatory order. B/W printer in good working order. Overall the system is in very good condition.
|
3000 - 5000 |
|
| 5 |
GE Medical Systems Logiq 5 Ultrasound
|
Serial number:- 53551SM8
Year of manufacture:- 2002
To include:- Sony UP_895MDW b/w printer - serial number 66473, GE 2294517 5.0Mhz probe - serial number 82887YM3, GE 2050357EV1 2.5Mhz probe - serial number 11401PD7, GE 2297883 intra vaginal probe - serial number 112806WX5, Panasonic AG-MD835E VCR - serial number L1TC00509, manual, presets and back up CD rom.
***Our Engineer’s comments:- Powers up and system self test/boot up ok. All controls and displays fully functional. Touch screen and menu navigation all in good working order. All three probes are working and in very good condition. All tested functions of available modes are in working order. VCR powers up ok. System is in very good working condition.
|
3000 - 5000 |
|
| 6 |
Eschmann Little Sister 5 Bench Top Autoclave
|
With vacuum
Serial number:- LSSA-1D-1824
Year of manufacture:- 2001
To include:- Eschmann X6 instruments, trays and a Eschmann tray handle
***Our Engineer’s comments:- LS5 powers up and self tests ok. Tested with two cycles with and without vacuum. Both sterilisation cycles completed without any faults or errors. The heating, vacuum and sterilisation phases are all in good working order. Displayed temperature and pressure are up to the required. LS5 without built in printer. Chamber, trays and door gasket are in very good condition. Steriliser is in very good condition and functional at the time of test.
|
500 - 800 |
|
| 7 |
Eschmann LS5 Bench Top Autoclave
|
With vacuum
Serial number:- LSSB-3C-1008
Year of manufacture:- 2003
To include:- Eschmann X3 instruments, trays and a Eschmann XI wire tray.
***Our Engineer’s comments:- LS5 powers up and displays ok, no errors produced. Tested in both vacuum and non-vacuum cycle, both in good working order. Temperature and pressure displays are within the norm for sterilisation of an open and wrapped load. Printer working ok. LS5 in excellent condition and in working order at the time of test.
|
500 - 800 |
|
| 8 |
Aloka Prosound SSD-5000 Ultrasound
|
Serial number:- M01000
Year of manufacture:- 2000
To include:- Aloka UST-9115-5 probe - serial number M01146, Aloka UST-9118 probe - serial number M00438, Aloka UST-9119 probe - serial number M01093, Sony UP-895CE printer and a Sony SVO-9500MDP VCR.
***Our Engineer’s comments:- Powers up and system self test ok. Keyboard and control console keys fully functional. Track ball and menu touch screen in good working condition. All mode selections - M, B, Dopple and flow in good working condition. B/W printer ok, monitor gives a good image. System is in good working order at the time of test. VCR - power functional.
|
2000 - 4000 |
|
| 9 |
Toshiba Tosbee Ultrasound
|
Serial number:- 45828
To include:- Toshiba PVF-375MT probe - serial number 04543339, Toshiba PVF-6205T probe - serial number 4592180 and a Toshiba TP-8700 printer.
***Our Engineer’s comments:- Powers up system self test ok. Keyboard and control console key in good working order. Gain, STC and depth range control ok. Track ball function mode B& M all working. Good display on monitor. Both probes are intact and fully functional. B/W printer in good working order. System is in very good condition and functioning at the time of test.
|
2000 - 4000 |
|
| 10 |
Alcon Laboratories Opthalmic Phaco Surgical System
|
Model:- Accurus 200PS
Serial number:- 0001499201X
Year of manufacture:- 05/2000
To include:- 7 x handpieces, 10 x boxes of accessories and a foot switch.
***Our Engineer’s comments:- Powers up and system self test ok. Display/LED test ok. Unit comes with utilities operator screen displayed. Touch screen fully operational selection of all operating modes possible. Pole height adjustment working ok. Both illuminator ports working. Remote control and foot switch in good working condition. No further test with hand pieces and accessories were conducted. System is in very good condition and functional at the time of test.
|
800 - 1200 |
|
| 11 |
Alcon Laboratories Opthalmic Phaco Surgical System
|
Model:- Accurus 200PS
Serial number:- 001808101XREVL
Year of manufacture:- 07/2000
To include:- 7 x handpieces, 10 x boxes of accessories and a foot switch.
***Our Engineer’s comments:- Powers up and system self test ok. Display/LED test ok. Unit comes with utilities operator screen displayed. Touch screen fully operational selection of all operating modes possible. Pole height adjustment working ok. Both illuminator ports working. Remote control and foot switch in good working condition. No further test with hand pieces and accessories were conducted. System is in very good condition and functional at the time of test.
|
800 - 1200 |
|
| 12 |
Siemens Siremobil Compact C-Arm
|
Serial number:- 024395C
Year of manufacture:- 12/1998
To include:- foot switch
***Our Engineer’s comments:- Powers up and system self test ok. Both workstation and c-arm control console working. Diaphragm and iris control are functional. Foot switch and exposure push button working ok. Keyboard and control console and display in good working order. Image storage and review ok. Exposure initiated and good quality image obtained. C-arm mechanical movement ok. c-arm height adjustment motorised and working ok. System is in very good working order at the time of test.
|
3000 - 5000 |
|
| 13 |
Coolview Wall Mounted Examination Lights x 2
|
***Our Engineer’s comments:- High intensity 50 watt examination light, 48,000 lux @ 0.5M. Comes with fixings. brand new still in box.
|
200 - 300 |
|
| 14 |
Coolview Desk Mounted Examination Lights x 2
|
***Our Engineer’s comments:- High intensity 50 watt examination light, 48,000 lux @ 0.5M. Comes with fixings. brand new still in box.
|
200 - 300 |
|
| 15 |
Coolview Mobile Examination Lights x 2
|
***Our Engineer’s comments:- High intensity 50 watt examination light, 48,000 lux @ 0.5M. Comes with fixings. brand new still in box.
|
200 - 300 |
|
| 16 |
Philips BV25 C-Arm X-Ray Unit
|
Serial number:- 28968200070
Year of manufacture:- 1995
To include:- Philips 45512668354 foot switch - serial number KFMS553700041, Philips 980760080001 cassette holder - serial number 2891420, Philips 98077530001 single monitor - serial number 29220801812, Kodak X-Omatic video film holder - 8X10inches, operator and service manual.
***Our Engineer’s comments:- Powers up and system self test ok. Display and control console in good working order. Test exposure in fluroscopy, pulse fluroscopy and radiography mode, all fully operational good quality image display and the single b/w monitor, shutters, collimator, field size and image rotation controls are in good working order. Foot switch and hand held exposure control working ok. system is in very good condition at the time of test. II = 6inches .
|
3000 - 5000 |
|
| 17 |
Philips BV25 C-Arm Mobile X-Ray Unit
|
Serial number:- 2691500002
Year of manufacture:- 1998
To include:- Philips 980775300014 single monitor - serial number 27058301212 and a Philips foot switch.
***Our Engineer’s comments:- Powers up and system self test ok. Display and control console in good working order. test exposure in fluroscopy, pulse fluroscopy and radiography mode, all fully opertional. Good quality image displayed on the single monitor. Shutter, collimator field size and image rotation control in good working condition. Foot switch and hand held exposure control working ok. Imager also powers up and system self test ok. System in good working condition at the time of test. Image intensifier flat surface, paint work has a few scratches. II - 6inches. C-arm swing brake needs slight adjustment.
|
3000 - 5000 |
|
| 18 |
Olympus Stack System
|
To include:- Olympus CLV-S40 light source - serial number 7303070, Olympus OTV-S6 video processor - serial number 7012598, Olympus OTV-S5 camera head - serial number AR-T12, Sony UP-2850P colour video printer - serial number 51642, Sony Trinitron colour monitor - serial number 2016420, Manual and a WM-SC workstation trolley.
***Our Engineer’s comments:- All systems power up ok. Light source comes on with brightness, produces an excellent image with camera head connected. White balance, gain and all controls working ok. Camera cable checked and found to be in good condition. Very good displayed image on colour monitor. Video printer needs a new ribbom and photo paper. System in very good condition at the time of test.
|
1000 - 2000 |
|
| 19 |
Zeiss S4 Operating Microscope
|
Serial number:- 281868
To include:- Zeiss foot switch - serial number 282707.
***Our Engineer’s comments:- Powers up ok, xenon and halogen light source is working ok. motorised x-y movement, focus and zoom in good working order. Operating distance = 170. Very good magnification and focussing - clear optical system. Zooming control only on backup OPMI motor supply switch / not possible on foot switch. All other functions are working ok.
|
1000 - 2000 |
|
| 20 |
Eschmann MR Operating Table
|
Serial number:- MR408
To include:- Eschmann cushion set
***Our Engineer’s comments:- Hydraulic lift and lowering in good working order, mechanical positioning fully functional, cushions in very good condition. Brakes and castor in working order. Operating table in very good condition and fully operational at the time of test.
|
800 - 1200 |
|
| 21 |
Eschmann LS5 Non Vacuum Table Top Steriliser
|
Serial number:- LSNA-IA-1592
To include:- manual, service & calibration record, and
3 x trays and a tray handle.
***Our Engineer’s coments:- Powers up ok. Cycle option display, all controls and displays are in good working order. Test cycle on 134degrees. Open instrument - no errors detected. Complete cycle and indicator for sterilisation - pass. All temperature and pressure within normal tolerance - see printout. printer in good working order. System working at the time of test. Updated service record.
|
600 - 1000 |
|
| 22 |
GE Medical Systems Stenoscop 2 C-Arm X-Ray Unit
|
Serial number:- NA50141105
To include:- GE Medical 8284900035 work station - serial number NA50141105, twin monitor and a GE Medical remote control.
Year of manufacture: 1995
***Our Engineer’s comments:- Powers up and system self test ok. Control console and display in good working order. Test exposure for pulse flurosopy, stand and fluroscopy, high intensity fluroscopy and radiography all fully functional. Satisfactory display obtained on the twin monitor. image storage/review, and transfer from monitor A to B fully functional. Shutters/collimator and image rotation in good working order. Foot switch and hand held exposure switch in good working condition. B/W thermal printer fully operational. Workstation keyboard and control in good working order. System in good working order when tested.
|
3000 - 5000 |
|
| 23 |
Toshiba Nemio 20 Diagnostic Ultrasound
|
Serial number:- E2513602
To include:- Toshiba PVM-651VT transvaginal probe - serial number A0612591, Mitsubishi P91 b/w printer, Toshiba PVM-375AT convex probe and a preset back-up CD Rom.
Year of manufacture: 2002
***Our Engineer’s comments:- System powers up and system self test ok. Probe self calibration ok, display and control console fully functional. Both transducers in good working order, 20, M, power doppler, colour doppler, pw modes in good working condition. Gain, track ball, freeze, soft key all functioning. B/W printer in good working order. VCR power up ok. System is in good working order and in good condition at the time of test.
|
3000 - 5000 |
|
| 24 |
Aloka SSD-1200 Echo Camera Diagnostic Ultrasound
|
Serial number:- 3IMI3584
To include:- Aloka 3.5Mhz ASt-9390-305 convex probe - Serial number 2510313, Aloka UST-936 7.5Mhz probe - serial number 9A3237, Aloka UST-556T 7.5Mhz probe - serial number 7A013148, Aloka UST-657-5 5Mhz probe - serial number 9A3043, Mitsubishi b/w printer and a service manual.
Year of manufacture: 1995
***Our Engineer’s comments:- System powers up and system self test ok. Probe self calibration ok, display and control console fully functional. Both transducers in good working order, 20, M, power doppler, colour doppler, pw modes in good working condition. Gain, track ball, freeze, soft key all functioning. B/W printer in good working order. VCR power up ok. System is in good working order and in good condition at the time of test.
|
2000 - 4000 |
|
| 25 |
ERBE ICC 200 Surgical Diathermy Unit
|
Serial number:- 8-1007
To include:- ERBE foot switch and a cut/coag
***Our Engineer’s comments:- Powers up and self test ok. Display and soft keys fully functional. Cut effect selector and power selector in good working order. Coag, forced, soft and bipolar selector working ok with the power selector. Cut/coag output check ok. Neutral plate mounting circuit in good working order. System functional and in satisfactory condition at the time of test. System mounted on trolley with shelf.
|
700 - 1000 |
|
| 26 |
Philips Practix 2000 Mobile X-Ray Unit
|
Serial number:- 96-37-397
Year of manufacture:- 1996
To include:- exposure switch and XI key.
***Our Engineer’s comments:- Powers up ok. battery charger in good working order. KV and MAS selector working ok. Exposure push switch functioning. Electromagnetic brake release for tube and assembly movement in good working order. Drive motor and control for unit in good condition. Exposure test ok - system in working condition at the time of test.
|
1000 - 2000 |
|
| 27 |
Zeiss Universal S3B Operating Microscope
|
Serial number:- 304903-9901
***Our Engineer’s comments:- Powers up ok. Both halogen lamps working. Fibre optic cable in good condition. joy stick control for motorised zoom and focus in good working order. Clear and high quality optical system. 12.5X eye piece, objective lens = 200mm. Manual x-y rotation movement in good working order. System in good condition at the time of test.
|
1000 - 2000 |
|
| 28 |
Maquet Operating Table
|
Model number:- 1117-OOAO
Serial number:- 00413
Year of manufacture:- 1994
***Our Engineer’s comments:- hydraulic lift in good working condition. Castors, and brakes are functional. All mechanical movement - head rest, back rest, foot rest (split) in excellent working order. Side way tilt and trendelenburg position also working. Table is in very good working condition.
|
800 - 1200 |
|
| 29 |
Olympus Evis Stack System
|
To include:- Sony trinitron colour monitor - serial number 2006738, olympus CV-240 video processor - serial number 7037481, Olympus CLV-U40 light source - serial number 7039930, Olympus MH-977 pig tail, Olympus MA-30 keyboard - serial number 7600120 and a Olympus WM-30 trolley.
***Our Engineer’s comments:- Powers up ok. Light source xenon lamp on, brightness control ok, air pump working. Processor fully operational, good image quality displayed, controls ok - keyboard functioning ok. System in very good condition at the time of test.
|
1000 - 2000 |
|
| 30 |
Zeiss KSK 150-FC Colposcope with Camera System
|
Serial number:- 320236
year of manufacture:- 2002
To include:- Sony DXC-LS1P camera control unit - serial number 401835, Sony CMA-DICE camera adapter unit - serial number 70362, Zeiss dust cover and video cables (BNC& S-Video).
Our Engineer’s comments:- powers up ok, lamp working, good light beam transmission, fibre optic cable in good condition. Light brightness control in good working order. Eye pieces lens 10x and objective lens = 300. Trolley and arms in good condition and movement. Camera controller and adapter give high quality image and image adjustments. system fully functional and in good condition at the time of test. System comes without monitor.
|
800 - 1200 |
|
| 31 |
Ohmeda Excel 210SE Anaesthetic Machine
|
Serial number:- AMAA04887
To include:- Ohmeda 7800 ventilator - serial number CAXA00147, HP M1094B anaesthesia monitoring system - serial number 3638171983, Agilent M1026A apnograph monitor and medical gas hoses.
***Our Engineer’s comments:- System power up and self test ok. Ventilator in good working order and tested with medical air and test lungs. Flow sensor and oxygen sensor working ok. the multiparameter monitor tested out ok. Full ECG trace and SPO2 signal, all in good working order. Capnograph detect CO2 and trace graph display indicates system in good working order at the time of test. No cuff for NIBP.
|
1000 - 2000 |
|
| 32 |
Olympus Stack System
|
To include:- Olympus OTV-S6H-10 camera head - serial number 7705755, Olympus camera head without lens - serial number 7800547, Olympus OTV-S6 video processor - serial number 7910856, Olympus CLV-S30 light source - serial number 7200351 and a Sony PVM-14M2MDE colour monitor - serial number 2002418.
***Our Engineer’s comments:- All units power up ok, light source lamp (xenon and halogen) in working condition. Brightness control ok. Video processor in good working condition. Both cameras are working ok, one without lens. Colour monitor is in good working order. System is in good working condition at the time of test.
|
1000 - 2000 |
|
| 34 |
Olympus Stack Trolley
|
To include:- Olympus CV240 video system centre - serial number 71169559, Olympus CLV-U40 light source - serial number 701392, Sony SVO-9500MDP VCR, Sony UP-890MD video printer and a Panasonic AG-7350 VCR.
***Our Engineer’s comments:- Light source and processor power up ok. Xenon lamp on, brightness control ok in automatic and manual mode. Video processor in good working order, good image quality processed. White balance, image enhance all in good working order. Both VCR’s power up ok. B/W printer is in good working condition. Keyboard fairly new and functional. System in good working order. System comes without monitor or pig tail lead.
|
1000 - 2000 |
|
| 35 |
Drager Babytherm 8000 Radiant Heater
|
Serial number:- ARJF-0028
To include:- gel mattress
***Our Engineer’s comments:- Powers up ok, radiant heater working ok and temperature setting maintained. Over head light also in good working order. Skin temperature mode in good working order, skin temp probe not included. Motorised table height adjustment working ok. Unit in good condition at the time of test.
|
500 - 1000 |
|
| 36 |
Drager Babytherm 8000 Radiant Heater
|
Serial number:- ARKM-0061
To include:- gel mattress
***Our Engineer’s comments:- Powers up ok, radiant heater working ok and temperature setting maintained. Over head light also in good working order. Skin temperature mode in good working order, skin temp probe not included. Motorised table height adjustment working ok. Unit in good condition at the time of test.
|
500 - 1000 |
|
| 37 |
Drager Babytherm 8000 Radiant Heater
|
Serial number:- ARJF-0031
To include:- gel mattress
***Our Engineer’s comments:- Powers up ok, radiant heater working ok and temperature setting maintained. Over head light also in good working order. Skin temperature mode in good working order, skin temp probe not included. Motorised table height adjustment working ok. Unit in good condition at the time of test.
|
500 - 1000 |
|
| 38 |
Zeiss EMD 523 Ceiling Mounted Operating Microscope
|
Serial number:- 237477
To include:- Zeiss foot switch, 2 x prismatic ALM ceiling mounted theatre lights and a Sony ceiling mounted monitor.
***Our Engineer’s comments:- Unit was fully functional when dismantled. In very good condition.
|
800 - 1200 |
|
| 39 |
Toshiba Power Vision 6000 Ultrasound
|
Serial number:- A1615228
Year of manufacture:- 2001
Version:- V4.01E2000
To include:- Toshiba PLW-1304AT probe - serial number 9980653743 and a Toshiba PVW-375AT probe - serial number D1612019.
***Our Engineer’s comments:- Powers up and system self test ok. Control console and display in good working order. Both transducers are in good working order. M, B, PW, CW, colour modes all in good order. Track ball, b/w printer and touch screen working ok. Touch screen with slight discolouration of the bottom. System in satisfactory working order and condition at the time of test.
|
2000 - 4000 |
|
| 40 |
Sonosite 180 Plus Ultrasound
|
Serial number:- 0300N6
Year of manufacture:- 2002
To include:- Sonosite L38/10 5Mhz probe - serial number 020869 and a Sonosite C15/4 2Mhz probe - serial number 0209KJ.
***Our Engineer’s comments:- Powers up and system self test ok. Control console and LCD screen in working order. Both transducers are in good condition and display good images. Battery and mains operation ok. All modes B, M, CPD and dopplers are in good working order. Gain and track ball are working ok. One side of limit lock broken, otherwise in good condition.
|
2000 - 3000 |
|
| 41 |
Topcon SL-5D Slit Lamp
|
Serial number:- 501339
To include:- camera, flash unit and tonometer.
***Our Engineer’s comments:- Powers up ok. Supply to lamp ok, lamp working with intensity selector switch ok. Good optical quality and magnification. Focusing very good, camera and flash unit also in good working order. Table weight adjustment working ok. System in good condition.
|
800 - 1200 |
|
| 42 |
Datex AS/3 Anaesthetic Machine
|
Serial number:- 4005760
To include:- Datex isoflurane vaporiser, M-ESTPT, M-NIBP and M-CAION modules.
***Our Engineer’s comments:- System powers up and self tests ok. Anaesthetic system pressure in good order. Medical gas supply and gauge pressure reading ok. Ventilator AS/3 fully functional, tested with test lung; setting, alarm limits, graphic waveform display and numeric display in satisfactory working order. Patient multiparameter monitor AS/3 display and gives good measurement for SPO2, ECG trace, CO2, spirometry and NIBP values. System is in good working order at the time of test. Monitor parameters:- NIBP, ECG, SPO2, temp, capnometry and spirometry.
|
1000 - 2000 |
|
| 43 |
Helion S Mobile Operating Light
|
Serial number:- 380-300-017-014
***Our Engineer’s comments:- Powers up ok. Halogen lamp working ok, good reflector with maximum light beam focussing. Mounted on castors and stand - new, very good condition. Autoclaveable handle.
|
500 - 1000 |
|
| 44 |
Helion S Mobile Operating Light
|
Serial number:- 380-300-017-014
***Our Engineer’s comments:- Powers up ok. Halogen lamp working ok, good reflector with maximum light beam focussing. Mounted on castors and stand - new, very good condition. Autoclaveable handle.
|
500 - 1000 |
|
| 45 |
Helion S Mobile Operating Light
|
Serial number:- 380-300-017-014
***Our Engineer’s comments:- Powers up ok. Halogen lamp working ok, good reflector with maximum light beam focussing. Mounted on castors and stand - new, very good condition. Autoclaveable handle.
|
500 - 1000 |
|
| 46 |
Berchtold Chromophade D500 Ceiling Mounted Operating Light
|
Serial number:- Z113N99-G10018
***Our Engineer’s comments:- Powers up ok, supply to lamp ok. Lamp in good working condition - comes with detachable handle, ceiling mounted plate and arm. Halogen lamp - 22.8V, ISOW x 2
|
300-500 |
|
| 47 |
Olympus Stack System
|
To include:- Olympus video system centre CV-240 - serial number 7047959, Olympus CLV-U40 light source - serial number 7010316, Olympus keyboard and a pigtail lead.
***Our Engineer’s comments:- Powers up ok. Light source xenon lamp on, brightness control working ok in manual and auto modes. Air pump functioning in all 3 settings. Video processor in good working condition, white balance, image enhance and all function keys working ok. Keyboard fully functional. All units inter-connected and mounted on Olympus WM-60 mobile workstation trolley with colour monitor. System is in very good condition and ready for use with pig tail MH-977.
|
1000 - 2000 |
|
| 48 |
Kaps SOM 22 Cold Light Wall Mounted Operating Microscope
|
Serial number:- 12838
***Our Engineer’s comments:- Powers up ok - light source working ok with both xenon reflector and lamp working ok. Fibre optic cable in very good condition. Good magnification and focussing. Clear and high quality optics. Objective lens = 200mm working distance. Comes with wall mounting plate / bracket. New system.
|
1000 - 2000 |
|
| 49 |
Astralite AL50M Mobile Examination Light
|
Serial number:- 17840/2
***Our Engineer’s comments:- Lamp powers up ok. Supply to lamp ok. Good and focussed illumination obtained. 22.8V, 50watt, halogen lamp, Mounted on stand with adjustable arm and castor.
|
400 - 600 |
|
| 50 |
Astralite AL50M Mobile Examination Light
|
Serial number:- 17840/1
***Our Engineer’s comments:- Lamp powers up ok. Supply to lamp ok. Good and focussed illumination obtained. 22.8V, 50watt, halogen lamp, Mounted on stand with adjustable arm and castor.
|
400 - 600 |
|